Centrifuga Avanti® J-26S
Beckman Coulter
The Avanti J-26S series is the only high performance centrifuge system you’ll ever need. Built upon benchmark performance, including high speed, low heat output, imbalance tolerance, and low noise the Avanti J-26S series has been redesigned for your ever-evolving lab and business needs. The Avanti J-26S is sustainable, easy-to-use, versatile, and safe.
High Performance
Application versatility: can process six liters of fermentor output in less than 10 minutes, deliver a subcellular fraction at 82,000 x g, run 24 microplates at a time, and separate live cell populations with elutriation.
Complete BioSafety* Systems
BioCertified** rotors, instruments and labware contain microbiological aerosols, while sterile filters prevent infectious materials from being expelled into the lab environment. Together, they enhance peace of mind for users and lab managers.
Dual System Rotor Recognition
Automatically detects which rotor is installed and prevents the rotor from running above its maximum-rated speed—without shutting down the run. All to ensure the safety of your lab.
Friction Reduction System
Enables quicker, quieter, more energy-efficient runs with better temperature control, which improves system reliability. High-torque, switched reluctance (SR) drive technology shortens cycle times.
Lightweight J-Lite Rotors
Consistently designed to be the most efficient rotors—for reduced run times and high g-force.
Ergonomic Design
Lower work surface for easier rotor handling. Foot pedal opens the door and improved door design swings up and out of the way.
Sustainability is Built In
Each instrument saves approximately 9.6 pounds of post-consumer materials from the landfill.
Datos Técnicos
UNSPSC: |
41103905 |
Height: |
86 cm (34 in); with door open 149 cm (58.5
in) |
Width: |
71 cm (28 in) |
Depth: |
86 cm (34 in); 102 cm (40.25 in) (including
air diverter extending from back panel) |
Weight: |
290 kg (640 lb.) |
Friction Reduction System (FRS): |
A mechanical rotary vane pump reduces chamber
pressure to 0.25 ATM (190 mm [7.5 in] Hg) |
Max Capacity Volume Range: |
6 liters |
Max Speed: |
26,000 rpm |
Maximum Heat Dissipation into Room under
Steady-State Conditions: |
2.0 kW (6,900 BTU/hr) |
Noise Level 0.91 meters (3 feet) in Front of
Centrifuge: |
57 dBA |
Refrigeration System: |
Non-CFC, non-ozone depleting refrigerant |
Rotor ID: |
Automatic |
Set Temperature: |
-10° C to +40° C (in 1° C increments) |
Set Time: |
1 Min - 3 Hours, Continuous Hold |
Speed Control: |
± 10 rpm of set speed or 0.1%, whichever is
greater |
Speed Range: |
100 to 26,000 rpm or equivalent RCF; 0 to
5,000 rpm for elutriation |
Temperature Control: |
± 2° C of set temperature (after
equilibration) |
Zonal Capability: |
Yes |
Single Tube Sample Volume Range: |
1.8 - 1,000 mL |
Ambient Temperature Range: |
16° C to 38° C |
Clearances Required: |
7.6 cm (3.0 in) sides; 16.0 cm (6.25 in)
rear |
Door: |
6.1 cm (2.4 in) thick plastic resin with steel
plate |
Humidity Restrictions: |
< 95% (noncondensing) |
Pollution Degree: |
2 |
Rotor Chamber Diameter: |
51.3 cm (20 in) |
Time -- Actual Display: |
Indicates run time remaining (timed run) or
elapsed time (hold run) |
Maximum Force at r max (g): |
82,000 x g |
Acceleration: |
Maximum or slow |
Deceleration: |
Maximum, slow, or
off |
|
|