UNSPSC: |
41103905 |
Height: |
86 cm (34 in); with door open 149 cm (58.7
in) |
Width: |
71 cm (28 in) |
Depth: |
86 cm (34 in); 102 cm (40.25 in) (including
air diverter extending from back panel) |
Weight: |
290 kg (640 lb.) |
Dimensions: |
865 mm x 876 mm X 711 mm (34.1 in x 34.5 in. x
28 in.) |
Approximate Acceleration Time: |
Maximum, slow, or timed (1 to 10 min from 0 to
500 rpm) |
Approximate Deceleration Time: |
Maximum, slow, timed (1 to 10 min from 500 to
0 rpm), or off |
Drive Cooling: |
Air Cooled |
Drive Type: |
SR* Drive |
Friction Reduction System (FRS): |
A mechanical rotary vane pump reduces chamber
pressure to 0.25 ATM (190 mm [7.5 in] Hg) |
Max Capacity Volume Range: |
6 liters |
Max Speed: |
26,000 rpm |
Maximum Heat Dissipation into Room under
Steady-State Conditions: |
2.0 kW (6,900 Btu/hr) |
Noise Level 0.91 meters (3 feet) in Front of
Centrifuge: |
< 57 dbA |
Refrigeration System: |
Non-CFC, non-ozone depleting refrigerant |
Rotor ID: |
Manual |
Set Temperature: |
-10° C to +40° C (in 1° C increments) |
Set Time: |
1 Min - 3 Hours, Hold |
Speed Control: |
± rpm of set speed or 0.1%, whichever is
greater |
Speed Range: |
100 to 26,000 rpm or equivalent RCF; 0 to
5,000 rpm for elutriation |
Temperature Control: |
± 2° C of set temperature (after
equilibration) |
Zonal Capability: |
No |
Single Tube Sample Volume Range: |
1.5 - 1,000 mL |
Ambient Temperature Range: |
15° C to 40° C |
Clearances Required: |
7.6 cm (3.0 in) sides; 16.0 cm (6.25 in)
rear |
Door: |
6.1 cm (2.4 in) thick structural foam with
steel plate |
Humidity Restrictions: |
< 95% (noncondensing) |
Pollution Degree: |
2 |
Rotor Chamber Diameter: |
51.3 cm (20 in) |
Time -- Actual Display: |
Indicates run time remaining (timed run) or
elapsed time (hold run) |
Maximum Force at r max (g): |
82,000 x g |